Anaesthesia Machine Checklist

βœ… ASA Summary of Anesthesia Machine Checkout Recommendations

πŸ”„ To Be Completed Daily

1. Verify that auxiliary oxygen cylinder and self-inflating manual ventilation device are available and functioning

πŸ‘₯ Provider and Technician

2. Verify that patient suction is adequate to clear the airway

πŸ‘₯ Provider and Technician

3. Turn on anesthesia delivery system and confirm that AC power is available

πŸ‘€ Provider or Technician

4. Verify availability of required monitors, including alarms

πŸ‘€ Provider or Technician

5. Verify that pressure is adequate on the spare oxygen cylinder mounted on the anesthesia machine

πŸ‘₯ Provider and Technician

6. Verify that the piped gas pressures are β‰₯ 50 psig

πŸ‘₯ Provider and Technician

7. Verify that vaporizers are filled and, if applicable, filler ports are tightly closed

πŸ‘€ Provider or Technician

8. Verify that there are no leaks in gas supply lines between flowmeters and common gas outlet

πŸ‘€ Provider or Technician

9. Test scavenging system function

πŸ‘€ Provider or Technician

10. Calibrate or verify calibration of the oxygen monitor, and check the low oxygen alarm

πŸ‘€ Provider or Technician

11. Verify that carbon dioxide absorbent is not exhausted

πŸ‘€ Provider or Technician

12. Perform breathing system pressure and leak testing

πŸ‘₯ Provider and Technician

13. Verify that gas flows properly through the breathing circuit during both inspiration and exhalation

πŸ‘₯ Provider and Technician

14. Document completion of checkout procedures

πŸ‘₯ Provider and Technician

15. Confirm ventilator settings and evaluate readiness to deliver anesthesia care (anesthesia time out)

πŸ‘€ Provider

πŸ•’ To Be Completed Prior to Each Procedure

2. Verify that patient suction is adequate to clear the airway

πŸ‘₯ Provider and Technician

4. Verify availability of required monitors, including alarms

πŸ‘€ Provider or Technician

7. Verify that vaporizers are filled and filler ports are closed

πŸ‘€ Provider

11. Verify that carbon dioxide absorbent is not exhausted

πŸ‘€ Provider or Technician

12. Perform breathing system pressure and leak testing

πŸ‘₯ Provider and Technician

13. Verify gas flows properly through breathing circuit (inspiration and exhalation)

πŸ‘₯ Provider and Technician

14. Document completion of checkout procedures

πŸ‘₯ Provider and Technician

15. Confirm ventilator settings and evaluate readiness to deliver anesthesia care (anesthesia time out)

πŸ‘€ Provider

Legend:

πŸ‘₯ = Provider and Technician

πŸ‘€ = Provider or Technician / Provider only